Volume 22, Issue 1 , Pages 25-28, January 2006
Development of head docking device for linac-based radiosurgery with a Neptun 10 PC linac
Abstract
Stereotactic radiosurgery is a method for focused irradiation of intracranial lesions. Linac-based radiosurgery is currently performed by two techniques: couch mounted and pedestal mounted. In the first technique a device is required to affix the patient's head to the couch and neoreover to translate it accurately. Structure of such a device constructed by the authors plus acceptance test performed for evaluation is described in the article.
A head docking device has been designed and constructed according to geometry of linac's couch and also desired functions. The device is cornpletely made from aluminum and consists of four major components: attachment bar, lower structure with four moveing accuracy mechanical stability and isocentric accuracy were assessed in the frame of acceptance test.
Translating accuracy, mechanical stability and isocentric accuracy were found to be respectively: 1 mm, 1.64 mm and 3.2 mm with accuracy of 95%.
According to AAPM report no. 54, a head docking device should translate head with an accuracy of 1 mm; this recommendation has been met. Moreover, we have demonstrated that the isocentric accuracy and mechanical stability of the device are sufficient that the device on confidently be used in stereotactic treatment.
Stereotactic radiosurgery, Head docking device, Translating accuracy, Mechanical stability
No full text is available. To read the body of this article, please view the PDF online.
To access this article, please choose from the options below
PII: S1120-1797(06)80007-4
doi:10.1016/S1120-1797(06)80007-4
© 2006 Associazione Italiana di Fisica Medica. Published by Elsevier Inc. All rights reserved.
Volume 22, Issue 1 , Pages 25-28, January 2006