Impact of high energy resolution detectors on the performance of a PET system dedicated to breast cancer imaging

  • Craig S. Levin

      Affiliations

    • Corresponding Author InformationAddress for correspondence: Craig S. Levin, Ph.D., Stanford University School of Medicine, 300 Pasteur Drive, MC 5128 Alway Building, Room M001 Stanford, CA 94305-5128, Phone: 650-736-7211, Fax: 650-724-1499.
  • ,
  • Angela M.K. Foudray
  • ,
  • Frezghi Habte

Abstract 

We are developing a high resolution, high sensitivity PET camera dedicated to breast cancer imaging. We are studying two novel detector technologies for this imaging system: a scintillation detector comprising layers of small lutetium oxyorthosilicate (LSO) crystals coupled to new position sensitive avalanche photodiodes (PSAPDs), and a pure semiconductor detector comprising cadmium zinc telluride (CZT) crystal slabs with thin anode and cathode strips deposited in orthogonal directions on either side of each slab. Both detectors achieve 1 mm spatial resolution with 3–5 mm directly measured photon interaction depth resolution, which promotes uniform reconstructed spatial resolution throughout a compact, breast-size field of view. Both detector types also achieve outstanding energy resolution (<3% and <12%, respectively for LSO-PSAPD and CZT at 511 keV). This paper studies the effects that this excellent energy resolution has on the expected system performance. Results indicate the importance that high energy resolution and narrow energy window settings have in reducing background random as well as scatter coincidences without compromising statistical quality of the dedicated breast PET data. Simulations predict that using either detector type the excellent performance and novel arrangement of these detectors proposed for the system facilitate ∼20% instrument sensitivity at the system center and a peak noise-equivalent count rate of >4 kcps for 200 microCi in a simulated breast phantom.

Breast cancer imaging, PET, detectors, cadmium zine telluride, scintillation detectors, position sensitive avalanche photodiodes

No full text is available. To read the body of this article, please view the PDF online.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1120-1797(06)80020-7

doi:10.1016/S1120-1797(06)80020-7

Physica Medica: European Journal of Medical Physics
Volume 21, Supplement 1 , Pages 28-34, 2006