| | Radiation dose to patients and staff during angiography of the lower limbs. Derivation of local dose reference levelsReceived 14 November 2007; received in revised form 26 February 2008; accepted 27 February 2008. published online 07 April 2008. Abstract BackgroundThe Euratom directive 97/43 recommends the use of patient dose surveys in diagnostic radiology and the establishment of reference dose levels (DRLs). PurposeTo perform measurements of the dose delivered during diagnostic angiography of the lower limbs using thermoluminescence dosimeters (TLDs), extraction of DRLs and estimation of the effective dose and radiation risk for this particular examination. MethodsDose measurement was performed on 30 patients by using TLD sachets attached in 5 different positions not only on the patient, but also to the radiologist. All the appropriate factors were recorded. Measurement of the ESD was performed after each examination. ConclusionRadiation dose variation depends on the physical characteristics of the patient, on the procedure preferences by radiologists and the difficulties in conducting procedures. The main reason for the increased patient dose, compared to other studies, is the number of frames rather than the duration of fluoroscopy. For DSA of the lower limbs, the DRL was chosen to be an entrance skin dose of 96.4 mGy in the pelvic region. The dose to the radiologist is negligible. Introduction  Digital Subtraction Angiography (DSA) is the golden standard technique for the diagnosis of vascular diseases [1], [2]. The main indications for DSA of the lower limbs are: atherosclerotic vascular disease, emboli, occlusive disease, and thrombosis, vascular trauma, preoperative planning and postoperative evaluation for reconstructive surgery, and other primary vascular abnormalities, including vascular malformations, vasculitis, entrapment syndrome, thoracic outlet syndrome, etc [3]. Due to the complexity of the procedure and the different cases, i.e. for patients with occlusive disease and thrombosis, with stents placed at the femoral arteries etc, the time that the pelvic area is irradiated can be long and therefore the dose to the patient and the personnel can be significant. However, if practice and protection are optimised, the dose to the patient will be as low as is compatible with the medical purpose [4]. The Euratom directive 97/43 [5], recommends the use of patient dose surveys in diagnostic radiology. The purpose of this study was to establish reference dose levels which can be applied to optimize the use of radiation. Especially in the field of therapeutic or diagnostic interventional radiology, where the fluoroscopy time is long, the number of the acquired images is large and the repetitions of the examinations is frequent, the need for the establishment of Dose Reference Levels (DRLs) is imperative. During a DSA examination the radiologists performing the examination are exposed to high levels of both scatter radiation from the patient and leakage radiation from the X-ray housing [4]. Generally speaking, the dose per examination is low but nevertheless, the accumulated dose might become significant over working years [6]. In vivo dosimetry of patients undergoing DSA is usually performed using two types of dosimeters: Dose–Area Product (DAP) meters [7], [8], [9], [10], [11], [12], and Thermoluminescent Dosimeters (TLD) [9]. Limitations in the use of TLDs for in vivo dosimetry, such as the number of dosimeters and the selection of the measurement location, can be overcome with the use of the maximum number of available dosimeters at standard locations. Nevertheless, in this specific study, TL dosimetry was prefered (rather than DAP meter dosimtery), since multiple projections are used and is difficult to measure the ESD of a specific point with the DAP meter. Thus, DAP values were not measured and evaluated at all for this particular group of patients. This study intends to evaluate, the radiation dose to the patient and staff during DSA of the lower limbs at Radiology Department, University Hospital of Larissa (UHL), Greece, using TLDs. In addition, it is aimed to estimate the risk of the radiation dose, to compare the results with the ones quoted in literature and to propose a local DRL for this investigation. Materials and methods  Investigation procedure DSA of the lower limbs was performed according to a standard protocol, by two experienced radiologists and four training radiologists. The procedure was performed utilizing Philips Integris 3000, an integrated X-ray system for all interventional procedures. The area selected for the puncturing was primarily the right femoral artery, but when this was impossible, the left artery was chosen. Initially, the catheter, fluoroscopically guided, was placed at the level of the renal arteries and after the administration of contrast medium digital images were acquired usually with a frame rate of two images per second (cine mode). Rarely, and if deemed necessary, a higher frame rate was selected (four images per second). Then, the X-ray tube was moved towards the feet in order to visualize all the arteries, usually acquiring one image per second. Whenever necessary, extra images were acquired with a tilt of the X-ray tube over the region of interest (two cases) or images were acquired for one of the two feet. The TLDs were calibrated under reproducible reference conditions using the same X-ray machine against an ionisation chamber model 9060/10X5-60 connected to a Radiation Monitor Controller model 9010 (Radcal Corporation, Monrovia, CA, USA). Both the chamber and the electrometer were calibrated for the energy range of 30–120 kV at the National Standard Laboratory (National Metrology Institute of the Greek Atomic Energy Commission). Calibration cycle of the TLDs was carried out every month in order to extract individual calibration factors. Patient dose measurements A total of 30 patients participated in this study. Indications included diagnosis and follow up of vascular diseases. The study was approved by ethics and research committee at UHL. Groups of four TLDs were packed into sachets. Eight sachets were positioned corresponding to five different exposure areas. An overall statistical uncertainty associated with the measurements of the TLDs was estimated to be 8%. All sachets were attached at the back of the patient in order to measure ESD. One sachet was placed at the level of aorta bifurcation, one at the pelvic area and 15 cm lower from the first one, one at the middle of the each femur, one at the back side of each knee and one at the back side of each foot. These positions were selected as they correspond to the positions where dynamic frames were acquired. ESD was measured as a total dose from both fluoroscopic and radiographic exposures. For each patient all the exposure parameters (i.e. radiographic data: kV, mAs, number of frames and region, fluoroscopic data: kV, mA and total screening time) and patient data (age, weight, height, clinical indication, and radiologist name) were recorded. Patient effective dose The ESD was used to calculate the effective dose. Estimation of the effective dose was done by the use of conversion factors provided by the National Radiological Protection Board (NRPB-SR 262) [13], [14]. The projections that were selected are: kidney PA for the TLDs that were at the level of the aorta bifurcation, pelvis PA for the TLDs that were at the level of the pelvis area, and shoulder PA for the rest [12]. The total effective dose was calculated by the addition of effective dose for each projection. Occupational dose measurements One sachet containing four TLDs, was attached outside the lead apron of 0.25-mm Pb equivalent thickness, at the chest. That position was selected, since the chest is the most common body part for dose evaluation using appropriate conversation factors [15]. Protective eyeglasses and thyroid collar (0.5-mm Pb equivalent thickness) were always worn. The staff radiation dose in diagnostic X-ray departments is routinely monitored by the official Greek authorities (Greek Atomic Energy Commission), with a dosimeter worn at the chest level, above the lead apron. Results  During a period of one year, dose monitoring was performed in 30 patients. The main indication for DSA of the lower limbs was atherosclerotic vascular disease, occlusive disease, and thrombosis. The mean age of the patients was 68.6 years, (range 48–90 years) and the mean height and weight were 170 cm and 76.8 kg, respectively. The mean screening time of fluoroscopy was 1.6 min (range 0.4–4.1 min) and the mean number of radiographic images was 103.7 (range 57–156). The mean value of kV for radiography was 62.3 kV (range 48–82 kV) and for fluoroscopy 69.5 kV (range 56–79 kV). The mean value of the mAs for radiography was 34.9 mAs (range 16.2–34.5 mAs) and the mean value of the tube current during fluoroscopy was 3.2 mA (range 1–5.9 mA). Table 1 presents the radiographic and fluoroscopic exposure factors for the five different regions of measurement. Table 2 presents the mean, range, and 75% of the ESD for the five positions of measurement and demonstrates that the most exposed places of the body are the abdomen and the pelvic region receiving 70.8 and 67.7 mGy, respectively. Also, the average ESD per image is presented. | | |  | | Abdomen | Pelvis | Femur | Knees | Feet | Average ESD per image |  |
|---|
 | ESD (mGy) | 70.8 | 67.7 | 24.3 | 18.4 | 9.7 | |  |  | 21.1–207.4 | 13.5–188.5 | 5.8–40.7 | 3.4–59.0 | 0.3–31.7 | 2.1 |  |  | 96.0 | 96.4 | 32.7 | 23.1 | 12.3 | |  | | | |
Table 3 presents the mean fluoroscopy time and the mean number of images acquired during angiography of the lower limbs, with corresponding results from other related studies. | a Femoral angiography. bangiography of the lower extremities. |
Table 4 represents the mean effective dose per procedure, which was 9.8 mSv, with corresponding results from other related studies. Fig. 1 presents the 75% value for pelvic ESD. The figure indicates that DRL for DSA of the lower limbs to be an ESD of 96.4 mGy in the pelvic region. Discussion  Patient exposure factors The mean fluoroscopic time during the procedure was 96 s (1.6 min) and the mean number of captured images was 103.7. Comparing these results with the ones from previous studies the screening time agrees with reported values, but the number of frames is quite large (Table 3). Thwaites et al. [8] reported that, on average, fluoroscopy contributed to 10% of the total radiation dose. Ruiz Cruces et al. [7] estimated the fluoroscopy contribution to be 36%. Bor et al. [9] calculated the contribution of fluoroscopy to be 21%. Mini et al. [16] reported that the contribution of the fluoroscopy to the DAP was around 25%. Struelens et al. [11] noted that fluoroscopy does not seem to be a substantial irradiating part of the procedure and that the dose to the patient depends mainly on the number of frames, the exposure parameters and the position of the radiation field. Hoskins et al. [17] estimated that digital spot film exposure contributed to 88% of the total effective dose. Thwaites et al. [8] reported the mean values of kV at the same areas of measurement during this study. The agreement, between corresponding areas, is good with a small deviation, of 10% maximum with the present study (Table 1). Ruiz Cruces et al. [7] reported average kV during radiography to be 66 kV. Struelens et al. [11] reported the results for eight different centres. The mean value of the results obtained during this study for the regions of abdomen and pelvis (71.25 kV) is very similar to the values of the present study with two exceptions, namely centre E and F1. In two centres, although they used similar exposure factors and number of frames, the effective dose is different (11.1–16.8 mSv for centre E and 7.7–8.3 mSv for centre F1). This difference may be due to the different fluoroscopy mode (different X-ray machine) or to the time of fluoroscopy. Bor et al. [9] reported average tube voltage 69 kV, and from a survey at four different centres [18] the tube voltage ranges from 67 to 88 kV. The above values demonstrate that fluoroscopy time gives a low contribution to the total dose of the patient and that the effective dose does not strongly depend on the tube voltage. Thus, it can be concluded that the calculated dose for the present study is relatively high, despite the short duration of fluoroscopy, mainly due to the large number of frames acquired. Patient effective doses, risk estimation The effective dose was derived from ESD using NRPB software [13], [14]. These factors were derived for a standard 70 kg anthropomorphic fantom irradiated with standard field sizes, and their use introduces an error in the estimation of the effective dose. The data for the present study (weight, field sizes) differ from those of the NRPB phantom, and doses to real patients may deviate significantly from the NRPB data even when the ESDs are equal. The effective doses derived in the present work are thus only indicative, compelled the research team for this study to consider this error of minor significance. The total effective dose is the sum of the effective dose at all the five regions. The regions contributing the most to the effective dose are the pelvic and the abdomen regions, while the smallest contributions stem from the femur, knees and feet. Struelens et al. [11] stated that 85% of the measured DAP stems from the exposure of the abdomen and pelvis. Bor et al. [9] assumed that 50% of the total DAP was contributed from the pelvic region. Hoskins et al. [17] estimated the contribution of torso exposure to be 98% of the total effective dose. From the results obtained during this study, it was calculated that the mean contribution of the abdomen and the pelvic regions to the effective dose to be 94.4%, (range 89.7–98.7%) of the total. The mean effective dose per procedure is 9.8 mSv, which is slightly higher than the mean effective dose reported in other studies. This can be explained by the large number of frames acquired, as the number of frames acquired contribute most to the effective dose. We must notice that the effective dose can be different than the reported one. This is not only due to the TLDs inaccuracy (8%) but also because there are places in the body that receiving dose and due to the luck of a dosimeter the dose is not recorded. Since the above happen mostly at the legs and their contribution at the total effective dose is minor we assume that the recorder dose is the total dose that the patient receives. It should be noticed that effective dose is not an additive quantity. The purpose of addition of each projection result is to be able to make simple comparison with the literature data which are usually given for the whole examination. Radiation risks estimation for fatal cancer per procedure was found to be 5.4 × 10−4 (540 per million) using the fatal cancer coefficient of 0.055 per Sv [19]. Staff doses Occupational dose is principally caused by scattered radiation and leakage from the X-ray tube housing. During the examination, the radiologist performing the procedure was protected by a movable ceiling mounted shield. No table mounted lead curtain exists in the department. The mean ESD at the chest level was recorded to be 0.15 mGy. The dose to the staff dose is very low under protective shield, since the radiologist wore lead apron of 0.25 mm lead equivalent with transmission varying from 4.3 to 10.2% at 70 kVp and thyroid collar of 0.5 mm lead equivalent with transmission varying from 0.6 to 1.6% at 70 kVp [20]. Taking into account the ESD at the chest level and assuming a percentage of 10% transmission the effective dose was calculated to be 0.023 mSv per procedure. Using the conversion factors obtained from International Commission of Radiological Protection (ICRP) [6], the eye lens dose and thyroid dose was estimated to be 14.92, and 0.75 μSv, respectively. Wearing thyroid collar and eyeglass reduces the dose to the radiologist to a negligible level. In general, the radiation dose of the staff is proportional to the dose to the patient. Scatter radiation could be decreased by decreasing the fluoroscopy time and the field size but also with the installation of a curtain shielding that would decrease the radiation that reach doctors extremities. References  [1]. [1]Angeli E, Vanzulli A, Venturini M, Biondi Zoccai G, Del Maschio A. J Nephrol. 2001;14:514–524. MEDLINE [2]. [2]Niazi K, Khan TH, Kirk A. Easley. Diagnostic utility of the two methods of ankle brachial index in the detection of peripheral arterial disease of lower extremities. Catheter Cardiovasc Interv. 2006;68(5):788–792. MEDLINE |
CrossRef
[3]. [3]Singh JH, Cardella JF, Cole PE, Grassi CJ, McCowan TC, Swan TL, et al. Quality improvement guidelines for diagnostic arteriography. J Vasc Interv Radiol. 2003;14:S283–S288. Full Text |
Full-Text PDF (75 KB)
[4]. [4]Williams JR. The interdependence of staff and patient doses in interventional radiology. Br J Radiol. 1997;70:498–503. MEDLINE [5]. [5]European Commission . Council directive of 30 June 1997 (97/43/Euratom) on health protection of individuals against the dangers of ionizing radiation in relation to medical exposure. Off J Eur Communities. 1997;. [6]. [6]International Commission of Radiological Protection . Recommendations of the international commission of radiological protection. 1990 Recommendations of the international commission on radiological protection. Ann ICRP. 1991;21–23. [7]. [7]Ruiz Cruces R, Garcia-Granados J, Diaz Romero FJ, Hernandez Armas J. Estimation of effective dose in some digital angiographic and interventional procedures. Br J Radiol. 1998;71:42–47. MEDLINE [8]. [8]Thwaites JH, Rafferty MW, Gray N, Black J, Stock B. A patient dose survey for femoral arteriogram diagnostic radiographic examinations using a dose–area produce meter. Phys Med Biol. 1996;41:899–907. MEDLINE |
CrossRef
[9]. [9]Bor D, Sancak T, Olgar T, Elcim Y, Adanali A, Sanlidilek U, et al. Comparison of effective doses obtained from dose–area product and air kerma measurements in interventional radiology. Br J Radiol. 2004;77:315–322. MEDLINE |
CrossRef
[10]. [10]McParland BJ. A study of patient radiation doses in interventional radiological procedures. Br J Radiol. 1998;71:175–185. MEDLINE [11]. [11]Struelens L, Vanhavere F, Bosmans H, Van Loon R, Geukens M. Data analysis from a multi-center, comparative study of angiographic examinations leading to practical guidelines for the optimisation of patient dose. Radiat Prot Dosimetry. 2005;117:87–92. MEDLINE |
CrossRef
[12]. [12]Le Heron JC. Estimation of effective dose to the patient during medical X-ray examinations from measurements of the dose–area product. Phys. Med. Biol. 1992;37(11):2117–2126. MEDLINE [13]. [13]Hart D, Jones DG, Wall BF. Normalised organ doses for medical X-ray examinations calculated using Monte Carlo Techniques. London: HMSO; 1994;. [14]. [14]Hart D. Authors' replay. Br J Radiol. 1998;71:995. [15]. [15]National council on Radiation Protection and measurements . Implementation of the principle of as low as reasonably achievable (ALARA) for medical and dental personnel. 1990;. [16]. [16]Mini RL, Schmid B, Schneeberger P, Vock P. Dose–area product measurements during angiographic X ray procedures. Radiat Prot Dosimetry. 1998;80:145–148. [17]. [17]Hoskins PR, Gillespie I, Ireland HM. Patient dose measurements from femoral angiography. Br J Radiol. 1996;69:1159–1164. MEDLINE |
CrossRef
[18]. [18]Bor D, Toklu T, Olgar T, Sancak T, Cekirge S, Onal B, et al. Variations of patient doses in interventional examinations at different angiographic units. Cardiovasc Intervent Radiol. 2006;29:797–806. MEDLINE |
CrossRef
[19]. [19]International Commission of Radiological Protection . Recommendations of the international commission of radiological protection. Biological and epidemiological information on health risks attributable to ionising radiation: a summary of judgements for the purposes of radiological protection of humans. <http://www.icrp.org/Health_risks.pdf>. [20]. [20]Christodoulou EG, Goodsitt MM, Larson SC, Darner KL, Satti J, Ping Chan H. Evaluation of the transmitted exposure through lead equivalent aprons used in a radiology department, including the contribution from backscatter. Med Phys. 2003;30:1033–1038. MEDLINE |
CrossRef
[21]. [21]Olerud HM, Saxebøl G. Diagnostic radiology in Norway from 1983 to 1993 – examination frequency and collective dose to patients. Radiat Prot Dosimetry. 1997;74:247–260. [22]. [22]Zoetelief J, Geleijns J, Kicken PJH, Thijssen MAO, van Unnik JG. Diagnostic reference levels derived from recent surveys on patients dose for various types of radiological examination in the Netherlands. Radiat Prot Dosimetry. 1998;80:109–114. [23]. [23]Wall BF. Diagnostic reference levels in the X-ray department. Eur Radiol Syllabus. 2004;14:66–73. [24]. [24]Verdun FR, Aroua A, Trueb PhR, Vock P, Valley JF. Diagnostic and interventional radiology: a strategy to introduce reference dose level taking into account the national practice. Radiat Prot Dosimetry. 2005;114:188–191. MEDLINE |
CrossRef
[25]. [25]Castellano IA, McNeil JG, Thorp NC. Assessment of organ radiation doses and associated risk for digital bifemoral angiography. Br J Radiol. 1995;68:502–507. MEDLINE |
CrossRef
a Medical Physics Department, University Hospital of Larissa, P.O. Box 1425, Larissa 41110, Hellas, Greece b Radiology Department, University Hospital of Larissa, P.O. Box 1425, Larissa 41110, Hellas, Greece Corresponding author. Tel.: +30 2410 682057; fax: +30 2410 670117.
PII: S1120-1797(08)00043-4 doi:10.1016/j.ejmp.2008.02.005 © 2009 Associazione Italiana di Fisica Medica. Published by Elsevier Inc. All rights reserved. | |
|