Physica Medica: European Journal of Medical Physics
Volume 26, Issue 3 , Pages 173-176, July 2010

Using the “STEP Test” to evaluate processing speed in Moroccan hospitals

  • K. Nfaoiu

      Affiliations

    • Faculte des Sciences, Université Mohammed V-Agdal, Rabat, Morocco
  • ,
  • F. Bentayeb

      Affiliations

    • Faculte des Sciences, Université Mohammed V-Agdal, Rabat, Morocco
    • Corresponding Author InformationCorresponding author. Tel.: +212 6 68 46 00 12.
  • ,
  • O. Basraoui

      Affiliations

    • Faculte des Sciences, Université Mohammed V-Agdal, Rabat, Morocco
  • ,
  • A.C.P. Azevedo

      Affiliations

    • Secretaria de Saude e Defesa Civil, Sup. de Vigilancia Sanitaria, Rio de Janeiro, Brazil

Received 8 May 2009; received in revised form 6 November 2009; accepted 20 November 2009. published online 07 December 2009.

Abstract 

The STEP Test is a simple method that can be used to quickly evaluate processing speed of automatic processors. It allows the identification of large deviations (>20%) in the processor-chemicals-film system. STEP means Sensitometric Test for the Evaluation of Processing. The performance of the processors taking part in the survey are calculated and compared to a standard processor where the film is processed under ideal conditions, according to manufacturer's recommendations. For each processor, in the dark room, a film is exposed to a sensitometer light, processed and the sensitometric curve is obtained. This survey was performed in 8 Moroccan hospitals in the cities of Casablanca and Rabat and included 11 processors. It was concluded that 50% of the processors present processing problems and are functioning under inadequate conditions since they present deviations larger than 20%.

Keywords: Automatic processors, Radiographic films, Sensitometry, STEP Test

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1120-1797(09)00067-2

doi:10.1016/j.ejmp.2009.11.002

Physica Medica: European Journal of Medical Physics
Volume 26, Issue 3 , Pages 173-176, July 2010