Volume 27, Issue 1 , Pages 52-57, January 2011
Physical characterization and comparison of two commercially available micro-MLCs
Article Outline
- Abstract
- Introduction
- Materials and methods
- Results and discussions
- Conclusions
- Disclaimer
- References
- Copyright
Abstract
In this study, the physical characteristics (penumbra width variation with the source size and shape, interleaf leakage, transmission through the leaves, and the tongue-and-groove effect) of two linear accelerators (BrainLAB's Novalis and Elekta's Synergy-S Beam Modulator) have been investigated. For similar square fields (about 4.5
cm
×
4.5
cm) with source-to-surface/skin-distance (SSD) ranging from 90
cm to 115
cm and measurements taken at the depth of Dmax
=
1.5
cm for 6
MV photon beam. The Novalis MLC has penumbra width of 2.4
±
0.11
mm–2.8
±
0.11
mm at the leaf-end and 2.2
±
0.1
mm–2.7
±
0.1
mm on the leaf-side; and those for the Synergy-S MLC are 4.4
±
0.17
mm–5.2
±
0.2
mm and 3.0
±
0.12
mm–3.5
±
0.12
mm. Upon rotating the Synergy-S collimator by 90° (i.e., shifting the leaf movement to the gun–target direction), significant reduction of the leaf-end penumbra width (17%) and increase of leaf-side penumbra width (28%) suggest an elliptical shape of the radiation source spot. Similar rotation of the collimator yielded reduction of the penumbras on both leaf-end (34%) and leaf-side (28%) for Novalis, indicating that the Novalis has a more symmetric source size. For all the field sizes and settings, BrainLAB's Novalis μMLC produce a smaller penumbra for simple square fields compared to the Elekta's Synergy-S. However, this difference became less pronounced for leaf-side penumbra and also for circular fields. The tongue-and-groove effect of the Novalis (23
±
0.9%) is slightly smaller than that of the Synergy-S (25
±
1%); while the interleaf leakage and leakage directly through leaves for Synergy-S (1.6
±
0.07% & 0.9
±
0.04%) are lower than that of Novalis (2
±
0.08% & 1.3
±
0.05%).
Keywords: Micromulti-leaf collimator, Penumbra, Interleaf leakage, Tongue-and-groove effect, Focal spot, IMRT
Introduction
Most commonly used integrated MLC ranges from 10
mm to 3
mm width as projected to the machine isocenter (normally 100
cm source-to-axis (SAD)). A smaller MLC leaf width is a desirable characteristic for obtaining improved field conformality around irregularly shaped targets. There are various studies in the literature related to comparison of physical characteristics and dosimetric impact of different MLC [1], [2], [3], [4], [5], [6], [7], [8], [9], [10], [11], [12], [14], [15], [16]; interleaf transmission [1], optimum leaf width [2], basic applications [3], comparison of micro-MLC [4], [5], [9], [11], characterization [6], [7], [8], [10], [16] and dosimetric consideration [12], [14], [15].
Using EDR2 film (Kodak, Rochester, NY), Clark et al. [4] reported the penumbra characteristics of the BrainLAB's Novalis and Varian's Millenium MLC, which have central leaf widths of 3
mm and 5
mm, respectively. The 3
mm MLC showed 80%–20% penumbra width of 2.4
±
0.1
mm in both the directions perpendicular and parallel to the leaf motion. The 5
mm MLC showed comparable penumbra width of 2.5
±
0.1
mm in the direction perpendicular to leaf motion and 3.2
±
0.1
mm in the direction parallel to leaf motion. Chern et al. [5] reported that the 3
mm μMLC of BrainLAB provided better target conformity and greater normal tissue sparing than the 5
mm MLC in stereotactic radiosurgery using dynamic conformal arcs. García-Garduño et al. [8] have recently reported penumbra (2.26
±
0.11
mm along leaf side and 2.31
±
0.11
mm along lead end), transmission (0.93
±
0.05%), and leakage (1.08
±
0.08%) of m3-μMLC of BrainLAB for 6
MV photon beam study using EBT film.
Patel et al. [14] investigated the dosimetric characteristics (leaf bank setup, penumbra width, leaf positioning reproducibility, interleaf leakage and leaf transmission) of a production pilot mMLC (Elekta Beam Modulator™, Elekta Oncology Systems, Crawley, UK) having 4 mm leaf width at isocenter. The penumbra values for leaf ends were measured to be between 4.2 mm and 4.8 mm for various large rectangular fields.
In this paper, the physical characteristics of μMLC from two linacs, one is the BrainLAB's Novalis and the other is Elekta's Synergy-S Beam Modulator, have been evaluated. The Novalis has three types of leaves projecting to 3
mm, 4.5
mm and 5.5
mm at isocenter, whereas the Synergy-S uses the Elekta Beam Modulator MLC with a 4
mm leaf width at isocenter. The Novalis machine is in clinical use for a long time; whereas, the Synergy-S is relatively new in the clinic (about 4 years). However, so far the physical characterization of the clinical Synergy-S has not been extensively evaluated in clinical settings. Both the linacs can be used for radiosurgery applications that require small field shaping with tight tumor margin, where the beam conformality to the PTV and narrow penumbra for any configuration of the collimator is highly desirable. The difference in spot size and shape may have significant effects on the variation of penumbra at different configuration of the collimators for both these machines. This aspect has not been investigated in any of the previous studies. Thus, it is important to investigate the physical characteristics especially, penumbra width for different collimator orientations. Additionally, interleaf leakage, direct leakage through the leaves, and the tongue-and-groove effect for these two machines which have comparable MLCs have been studied.
Materials and methods
System overview
BrainLAB's Novalis linac has 26 leaf pairs mounted at a distance of about 44
cm from the radiation source, i.e., the focal spot. The maximum field size at the isocentric plane that can be used is 9.6
cm
×
10.2
cm. The leaf width (mutually perpendicular to both leaf movement direction and the beam central axis) at 100
cm is 3
mm for central 14 leaf pairs, 4.5
mm for the next 3 leaf pairs on each side of the central leaves and 5.5
mm for the remaining 3 leaf pairs on each side of the central leaves. The physical leaf thickness is 6.4
cm and the leaf tip is rounded. There are two sets of backup jaws that work in combination with the μMLC leaves for enhancing beam attenuation.
Elekta's Synergy-S Beam Modulator has a total of 40 pairs of leaves with a 4
mm width measured at 100
cm and the physical leaf thickness is 7.5
cm, and the leaf tip is rounded. The MLC is installed at a distance of 39
cm from the source of radiation. The maximum field at the isocentric plane is 21
cm
×
16
cm. In this study, both linacs have been characterized for 6
MV photon energy for same or similar field sizes. The Synergy-S Beam Modulator system does not use a backup jaw system.
Penumbra measurement
In this study, the physical penumbra characteristics of the Elekta's Synergy-S μMLC and the BrainLAB's Novalis μMLC were investigated by exposing Kodak EDR2 (Eastman Kodak Inc., Rochester, NY) films at different configurations, i.e., various field sizes and shapes (SSD
=
100
cm and depth
=
Dmax are kept constant), SSDs (depth
=
Dmax), and depths of measurement (constant SSD
=
100
cm). For evaluating the effects of SSD, a range of SSD (90
cm to 115
cm) was considered while the films were placed at the depth of Dmax
=
1.5
cm. The EDR2 film was calibrated by generating the Hurter and Driffield (H&D) curve, plotting dose vs. optical density following the guideline of the AAPM Task Group 69 [13]. Exposed films were scanned using VIDAR's DiagnosticPRO Advantage film digitizer (VIDAR Systems Corp., Herndon, VA). To investigate the effect of the shape of the source, i.e., focal spot size, two sets of measurements were taken by orienting the collimator at 0° and at 90° with the gantry at 0°. Square and circular fields ranging from 3
cm to 10
cm were used. The SSD
=
100
cm was used for studying the effect of field size and shape. The effects of measurement depth were investigated for a depth ranging from 1.5
cm to 15
cm while the SAD was kept constant at 100
cm. The effective penumbra information was expressed as lateral distance of 80%–20% isodose lines [4], [7], [8], [9], [14], which is considered more relevant for clinical applications. For all the film measurements three sets of films were exposed for each of the settings and 10 measurements at different location were taken on the irradiated films for calculating mean and standard deviation (SD). The SD was incorporated into the mean value to show the measurement uncertainties as mean
±
SD.
Tongue-and-groove effect
The tongue and grove (T&G) effects were measured by pulling in a set of leaves to block the radiation beam while approximately 100
cGy (100
MU) dose was delivered. Then this set of leaves was retracted and the alternate set of leaves was brought into the field to block the beam and another 100
MU was delivered. The T&G effects as well as the leakage measurement were performed using the same EDR2 films as mentioned above in Section Penumbra measurement.
Leakage measurement
The leakage measurements were performed by exposing radiographic films while the radiation beam was blocked using the MLC for both Novalis and Synergy-S. For Novalis, the jaws were parked just outside of the maximum field (9.6
cm
×
10.2
cm), which was covered by the MLC only. The films were exposed to about 5000
cGy (5000
MU) placing at 100
cm SAD through 1.5
cm solid water buildup. The percentage leakage was defined for characterizing the interleaf leakage and leakage through leaves.
Results and discussions
Penumbra
The results of the penumbra measurements for 6
MV photon beam are presented in Fig. 1 through Fig. 3. In Fig. 1a, the penumbras of Synergy-S and Novalis for both leaf-end (along the direction of motion of the leaves) and leaf-side (in the direction perpendicular to the motion of the leaves) for SSDs from 90
cm to 115
cm at constant depth
=
Dmax and square field of 4.8
cm2 for Synergy-S and 4.2
cm2 for Novalis at isocenter of 100
cm, with the collimator at 0° orientation have been shown. It is to be noted that the square field of 4.2
cm2 for Novalis was created by using the central 14 pairs of leaves of 3
mm width, while the 4.8
cm2 field of Synergy-S was made by using 12 pairs of central leaves of 4
mm width. The leaf-end penumbra width for SSD 90–115
cm is in the range from 4.4
±
0.17
mm to 5.2
±
0.2
mm for Synergy-S (top curve) and it is much larger compared to the leaf-side penumbra (range 3.0
±
0.12
mm-3.5
±
0.12
mm) for this linac. On the contrary, both the leaf-end (range 2.4
±
0.11
mm–2.8
±
0.11
mm) and the leaf-side (range 2.2
±
0.1
mm–2.7
±
0.1
mm) penumbra widths for the Novalis (bottom two curves) are close to each other as well as much smaller than respective penumbras of the Synergy-S. The average difference of penumbra at the leaf-side and at the leaf-end is about 50% for the Elekta's Synergy-S Beam Modulator, while this difference is about 6% for the Novalis' μMLC. These differences between the Synergy-S and the Novalis penumbras could be mainly related to: (1) the source to diaphragm/collimator distance (SDD
=
44
cm for Novalis and SDD
=
39
cm for Synergy-S), (2) the source size and shape. Shorter SDD, and larger source size contribute to increase in the penumbra of Synergy-S (the contribution from field size difference of 0.6
cm, i.e., 4.8
cm vs. 4.2
cm is relatively small). The closeness of the leaf-end and leaf-side penumbras of the Novalis can be attributed to the shape and size of the focal spot (approximately circular, smaller) and the design of the leaf-end and the leaf-side (tongue-groove).

Figure 1
a) Penumbra width of Elekta's Synergy-S and BrainLAB's Novalis at Dmax
=
1.5
cm of 6
MV photon for SSD setup while the collimator was at 0-deg. (b) Penumbra of Elekta's Synergy-S and BrainLAB's Novalis at Dmax of 6
MV photon for SSD setup. Collimator at 90-deg.
The collimator system was rotated to the 90° position to determine if elongation of the radiation spot size explained the difference in the leaf end and leaf-side penumbra widths for the Elekta collimator. It was interesting to observe that after collimator rotation, the size of the two penumbra widths (the leaf-end and the leaf-side) for the Synergy-S moved closer to an average position relative to those seen in Fig. 1a (top two curves in Fig. 1b). The leaf-end penumbra decreased by 17% and leaf-side penumbra increased by 28% (see Table 1 below). This suggests that the shape of the X-ray spot is elliptical with the minor axis along the gun–target direct of the linac. On the other hand, similar rotation of the collimator yielded significant decrease of the penumbra widths on both leaf-end (34%) and leaf-side (28%) for Novalis (bottom two curves in Fig. 1b). This indicates that the Novalis has approximately circular source as compared to that of the Synergy-S.
Table 1. Change in penumbra width due to change in collimator orientation.
| Linac | Leaf-end average penumbra at Dmax for 90 | Leaf-side average penumbra at Dmax for 90 | ||||
|---|---|---|---|---|---|---|
| Collimator at 0-deg | Collimator at 90-deg | Change | Collimator at 0-deg | Collimator at 90-deg | Change | |
| Synergy-S | 4.8 | 4.0 | −17 % ↓ | 3.2 | 4.1 | 28 % ↑ |
| Novalis | 2.6 | 1.7 | −34 % ↓ | 2.5 | 1.8 | −28% ↓ |
The penumbras measured for different setups (field size and depth of measurement), are presented in Fig. 2. The left column (Fig. 2(a) for leaf-end and (b) for leaf-side) is from Synergy-S and the right column is from Novalis (Fig. 2c for leaf-end and 2d for leaf-side). For a constant SAD/SCD, the penumbra increase with the measurement depth is more pronounced for larger field sizes due to additional scattering.

Figure 2
Leaf-end and leaf-side penumbras for different square fields, depth, d
=
1.5, 5, 10, 15
cm for 6
MV photon for 100
cm SAD and collimator at 0-deg. (a,b) Elekta's Synergy-S, and (c,d) BrainLAB's Novalis.
The effective penumbras as measured for circular fields (from 3
cm to 9
cm in diameter) are comparable for Novalis' and Synergy-S' μMLC for larger field sizes and larger measurement depths (Fig. 3). The Novalis' μMLC has a consistently smaller penumbra for the circular fields and compared to the Synergy-S' MLC. It is to be noted that the measured penumbra for Novalis (BrainLAB) closely matched with that reported by Clark et al. [4], Chern et al. [5], and García-Garduño et al. [8]. The effective penumbras for these circular fields range from about 4
±
0.15
mm to 7
±
0.27
mm for the Synergy-S μMLC (Fig. 3a) and from about 3
±
0.12
mm to 6
±
0.23
mm for the BrainLAB's μMLC (Fig. 3b).

Figure 3
Penumbras for different circular fields, depth, d
=
1.5, 5, 10, 15
cm for 6
MV photon for 100
cm SAD and collimator at 0-deg. (a) Elekta's Synergy-S (on the left panel), and (b) BrainLAB's Novalis (on the right panel).
Tongue-and-groove (T&G) effect
The tongue-and-groove effects for 6
MV photon beam for Synergy-S and Novalis have been presented in Fig. 4. In fact leaves of the Synergy-S Beam Modulator do not have tongue and groove as such; they are machined at a particular angle with respect to the radiation beam, thus making the side of the leaf plane but blocking the beam by overlapping with the adjacent leaves. On the other hand, the Novalis is designed with conventional tongue and grove to minimized interleaf leakage. The maximum T&G effect for Synergy-S is about 25
±
1% and it is about 23
±
0.9% for Novalis (20
±
0.8% for 3
mm leaves and 23
±
0.9% for 5.5
mm leaves). It should be noticed that the smaller central leaves (3
mm width) produce smaller T&G effects as compared to 4.5
mm and 5.5
mm leaves (Fig. 4b).

Figure 4
Tongue-and-groove effects of the Elekta's Synergy-S (on the left panel) and BrainLAB's Novalis (on the right panel).
Leakage
In this study, the leakage for 6
MV beam has been quantified as percentage for a delivery of 5000
MU (about 5000
cGy). The average and maximum leakage for the Synergy-S MLC is 0.9
±
0.04% and 1.6
±
0.07%, respectively, (Fig. 5a). The average and maximum leakages of the Novalis are 1.3
±
0.05% and 2.0
±
0.08%, respectively. It is to be noted that the leakage through 3
mm leaves is larger than through 4.5
mm or 5.5
mm leaves (Fig. 5b). Thus, Synergy-S Beam Modulator has smaller leakage as compared to that of Novalis M3 μMLC, which is probably due to larger leaf width.

Figure 5
Intra- and interleaf leakage along the field center – (a) Elekta's Synergy-S (on the left panel), and (b) BrainLAB's Novalis (on the right panel).
Conclusions
In this study, the penumbra, tongue-and-groove effect, and transmission through leaves for the μMLC of two linacs (Elekta's Synergy-S and BrainLAB's Novalis) have been investigated. Measurements made using radiographic EDR2 films showed that both the BrainLAB's and the Elekta's μMLC produce comparable effective penumbras for circular fields for larger field sizes and larger measurement depths. The BrainLAB's μMLC has smaller effective penumbra for smaller circular fields (<6
cm in diameter), which are typical in radiosurgery and shallower (<5
cm) depths. However, for square fields, the Synergy-S appears to produce larger penumbra.
An important finding of this study is the large change of penumbra (17%–34%) while the collimator is rotated by 90° (Fig. 1 and Table 1). Appears that tighter penumbra can be achieved with collimator at 90° as compared to 0° configuration. This can be dosimetrically significant while planning with tighter margin to the tumor. The measurements with collimator rotation suggested the elliptical shape of the Synergy-S' radiation source and roughly circular shape of the Novails' radiation source. The elliptical shape of the sources has important implication on the size of the penumbra (especially for Synergy-S) and may be considered for improving dose distribution.
Disclaimer
The contents and comments of this manuscript are the sole responsibility of the authors and do not necessarily represent the views of the manufacturer and vendors of the systems involved.
References
- . Multileaf collimator interleaf transmission. Med. Phys. 1999;26:176–186
- . What is the optimum leaf width of a multileaf collimator?. Med. Phys. 2000;27:2494–2502
- Basic applications of multileaf collimators. AAPM Report No. 72 Madison, WI: Medical Physics Publishing; 2001;
- . Penumbra evaluation of the varian millennium and BrainLAB M3 multileaf collimators. Int. J. Radiat. Oncol. Biol. Phys. 2006;66:S71–S75
- . Is smaller better? Comparison of 3-mm and 5-mm leaf size for stereotactic radiosurgery: a dosimetric study. Int. J. Radiat. Oncol. Biol. Phys. 2006;66:S76–S81
- . Characterization of a multi-leaf collimator system. Int. J. Radiat. Oncol. Biol. Phys. 1993;25:181–192
- . A comparison of multileaf collimator system and alloy-block field shaping. Int. J. Radiat. Oncol. Biol. Phys. 1998;40:721–731
- . Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film. Appl. Clin. Med. Phys. 2008;9(3):
- . Dosimetric comparison of various multileaf collimators. Phys. Med. Biol. 2002;47:N159–N170
- . The design and performance characteristics of a multileaf collimator. Phys. Med. Biol. 1994;39:231–251
- . Comparison of a micro-multileaf collimator with a 5-mmleaf-width collimator for intracranial stereotactic radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2003;57:1443–1449
- . Dosimetric considerations for a multileaf collimator system. Med. Phys. 1996;23:1219–1224
- TG-69: radiographic film for megavoltage beam dosimetry. Med. Phys. 2007;34:2228–2258
- . Dosimetric characteristics of the Elekta Beam Modulator™. Phys. Med. Biol. 2005;50:5479–5492
- Stereotactic IMRT for prostate cancer: dosimetric impact of multileaf collimator leaf width in the treatment of prostate cancer with IMRT. J. Appl. Clin. Med. Phys. 2004;5:29–41
- Physical characteristics of a miniature multileaf collimator. Med. Phys. 1999;26:65–70
PII: S1120-1797(10)00007-4
doi:10.1016/j.ejmp.2010.01.005
© 2010 Associazione Italiana di Fisica Medica. Published by Elsevier Inc. All rights reserved.
Volume 27, Issue 1 , Pages 52-57, January 2011
