Physica Medica: European Journal of Medical Physics
Volume 27, Issue 2 , Pages 81-88 , April 2011

Practical use of Gafchromic® EBT films in electron beams for in-phantom dose distribution measurements and monitor units verification

  • Jad El Barouky

      Affiliations

    • Department of Medical Physics, Institut Curie, Paris, France
    • Corresponding Author InformationCorresponding author.
  • ,
  • Nathalie Fournier-Bidoz

      Affiliations

    • Department of Medical Physics, Institut Curie, Paris, France
  • ,
  • Alejandro Mazal

      Affiliations

    • Department of Medical Physics, Institut Curie, Paris, France
  • ,
  • Georges Fares

      Affiliations

    • Department of Medical Physics, Hôtel-Dieu de France, Beirut, Lebanon
  • ,
  • Jean-Claude Rosenwald

      Affiliations

    • Department of Medical Physics, Institut Curie, Paris, France

Received 13 November 2009 ,Revised 29 January 2010 ,Accepted 28 April 2010.

References 

  1. Devic S, Seuntjens J, Hegyi G, et al. Dosimetric properties of improved GafChromic films for seven different digitizers. Med Phys. 2004;31:2392–2401
  2. Chiu-Tsao ST, Duckworth T, Zhang C, et al. Dose response characteristics of new models of GAFCHROMIC films: dependence on densitometer light source and radiation energy. Med Phys. 2004;31:2501–2508
  3. Todorovic M, Fischer M, Cremers F, et al. Evaluation of GafChromic EBT prototype B for external beam dose verification. Med Phys. 2006;33:1321–1328
  4. Devic S, Tomic N, Pang Z, et al. Absorption spectroscopy of EBT model GAFCHROMIC film. Med Phys. 2007;34:112–118
  5. ISP, International Speciality Products. White Paper: Gafchromic® EBT. Self-developing film for radiotherapy dosimetry. New Jersey, USA; 2007.
  6. Fiandra C, Ricardi U, Ragona R, et al. Clinical use of EBT model Gafchromic film in radiotherapy. Med Phys. 2006;33:4314–4319
  7. Chiu-Tsao ST, Ho Y, Shankar R, et al. Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies. Med Phys. 2005;32:3350–3354
  8. Zeidan OA, Stephenson SA, Meeks SL, et al. Characterization and use of EBT radiochromic film for IMRT dose verification. Med Phys. 2006;33:4064–4072
  9. Butson MJ, Cheung T, Yu PK. Weak energy dependence of EBT gafchromic film dose response in the 50kVp–10MVp X-ray range. Appl Radiat Isot. 2006;64:60–62
  10. Cheung T, Butson MJ, Yu PK. Independence of calibration curves for EBT Gafchromic films of the size of high-energy X-ray fields. Appl Radiat Isot. 2006;64:1027–1030
  11. Cheung T, Butson MJ, Yu PK. Measurement of high energy X-ray beam penumbra with Gafchromic EBT radiochromic film. Med Phys. 2006;33:2912–2914
  12. Cheung T, Butson MJ, Yu PK. Post-irradiation colouration of Gafchromic EBT radiochromic film. Phys Med Biol. 2005;50:N281–285
  13. Sankar A, Ayyangar KM, Nehru RM, et al. Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification. Med Dosim. 2006;31:273–282
  14. Roland TF, Stathakis S, Ramer R, et al. Measurement and comparison of skin dose for prostate and head-and-neck patients treated on various IMRT delivery systems. Appl Radiat Isot. 2008;66:1844–1849
  15. Wilcox E, Daskalov G, Nedialkova L. Comparison of the epson expression 1680 flatbed and the vidar VXR-16 dosimetry PRO film scanners for use in IMRT dosimetry using gafchromic and radiographic film. Med Phys. 2007;34:41–48
  16. Mack A, Mack G, Weltz D, et al. High precision film dosimetry with GAFCHROMIC films for quality assurance especially when using small fields. Med Phys. 2003;30:2399–2409
  17. Aziz T, Budgell GJ, Mackay RI. Evaluation of Gafchromic EBT Film for Prostate IMRT Verification. Clin Oncol (R Coll Radiol). 2007;19:S24
  18. Su FC, Shi C, Papanikolaou N. Clinical application of GAFCHROMIC (R) EBT film for in vivo dose measurements of total body irradiation radiotherapy. Appl Radiat Isot. 2008;66:389–394
  19. Devic S, Seuntjens J, Abdel-Rahman W, et al. Accurate skin dose measurements using radiochromic film in clinical applications. Med Phys. 2006;33:1116–1124
  20. Cheung T, Butson MJ, Yu PK. Multilayer Gafchromic film detectors for breast skin dose determination in vivo. Phys Med Biol. 2002;47:N31–37
  21. Bilge H, Cakir A, Okutan M, et al. Surface dose measurements with GafChromic EBT film for 6 and 18MV photon beams. Phys Med. 2008;
  22. Yamauchi M, Tominaga T, Nakamura O, et al. GAFChromic film dosimetry with a flatbed color scanner for Leksell Gamma Knife therapy. Med Phys. 2004;31:1243–1248
  23. Wilcox EE, Daskalov GM. Evaluation of GAFCHROMIC EBT film for Cyberknife dosimetry. Med Phys. 2007;34:1967–1974
  24. Pantelis E, Antypas C, Petrokokkinos L, et al. Dosimetric characterization of CyberKnife radiosurgical photon beams using polymer gels. Med Phys. 2008;35:2312–2320
  25. Ciocca M, Orecchia R, Garibaldi C, et al. In vivo dosimetry using radiochromic films during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol. 2003;69:285–289
  26. Su FC, Liu Y, Stathakis S, et al. Dosimetry characteristics of GAFCHROMIC EBT film responding to therapeutic electron beams. Appl Radiat Isot. 2007;65:1187–1192
  27. Bufacchi A, Carosi A, Adorante N, et al. In vivo EBT radiochromic film dosimetry of electron beam for Total Skin Electron Therapy (TSET). Phys Med. 2007;23:67–72
  28. Reynard EP, Evans MD, Devic S, et al. Rotational total skin electron irradiation with a linear accelerator. J Appl Clin Med Phys. 2008;9:2793
  29. Fiandra C, Ragona R, Ricardi U, et al. Absolute and relative dose measurements with Gafchromic EBT film for high energy electron beams with different doses per pulse. Med Phys. 2008;35:5463–5470
  30. IAEA, International Atomic Energy Agency. Technical reports series No. 430: commissioning and quality assurance of computerized planning systems for radiation treatment of cancer. Vienna, Austria; 2004.
  31. Gillis S, De Wagter C, Bohsung J, et al. An inter-centre quality assurance network for IMRT verification: results of the ESTRO QUASIMODO project. Radiother Oncol. 2005;76:340–353
  32. Butson MJ, Yu PK, Cheung T, et al. High sensitivity radiochromic film dose comparisons. Phys Med Biol. 2002;47:N291–295
  33. Butson MJ, Cheung T, Yu PK. Absorption spectra variations of EBT radiochromic film from radiation exposure. Phys Med Biol. 2005;50:N135–140
  34. Fuss M, Sturtewagen E, De Wagter C, et al. Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance. Phys Med Biol. 2007;52:4211–4225
  35. Butson MJ, Cheung T, Yu PK. Scanning orientation effects on Gafchromic EBT film dosimetry. Australas Phys Eng Sci Med. 2006;29:281–284
  36. IAEA, International Atomic Energy Agency. Technical reports series No. 398: absorbed dose determination in external beam radiotherapy: an internal code of practice for dosimetry based on standards of absorbed dose to water. Vienna, Austria; 2000.
  37. (AAPM Task Group 70) , Gerbi BJ, Antolak JA, Deibel FC, Followill DS, Herman MG, et al. Recommendations for clinical electron beam dosimetry: supplement to the recommendations of Task Group 25. Med Phys. 2009;36(7):3239–3279
  38. van Battum LJ, Hoffmans D, Piersma H, et al. Accurate dosimetry with GafChromic EBT film of a 6MV photon beam in water: what level is achievable?. Med Phys. 2008;35:704–716
  39. Devic S, Seuntjens J, Sham E, et al. Precise radiochromic film dosimetry using a flat-bed document scanner. Med Phys. 2005;32:2245–2253
  40. Menegotti L, Delana A, Martignano A. Radiochromic film dosimetry with flatbed scanners: a fast and accurate method for dose calibration and uniformity correction with single film exposure. Med Phys. 2008;35:3078–3085
  41. Lynch BD, Kozelka J, Ranade MK, et al. Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT GAFCHROMIC film. Med Phys. 2006;33:4551–4556
  42. Paelinck L, De Neve W, De Wagter C. Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry. Phys Med Biol. 2007;52:231–242
  43. Dutreix J, Dutreix A. Comparative study of a series of ionization chambers within 20 and 10MeV electron fluxes. Biophysik. 1966;3:249–258
  44. Feldman A, de Almeida CE, Almond PR. Measurements of electron-beam energy with rapid-processed film. Med Phys. 1974;1:74–76
  45. Niroomand-Rad A, Gillin MT, Kline RW, et al. Film dosimetry of small electron beams for routine radiotherapy planning. Med Phys. 1986;13:416–421
  46. Niroomand-Rad A. Film dosimetry of small elongated electron beams for treatment planning. Med Phys. 1989;16:655–662
  47. Suchowerska N, Hoban P, Butson M, et al. Directional dependence in film dosimetry: radiographic and radiochromic film. Phys Med Biol. 2001;46:1391–1397

PII: S1120-1797(10)00025-6

doi: 10.1016/j.ejmp.2010.04.001

Physica Medica: European Journal of Medical Physics
Volume 27, Issue 2 , Pages 81-88 , April 2011