Volume 28, Issue 1 , Pages 7-12, January 2012

Patient dosimetry and a novel approach to establishing Diagnostic Reference Levels in dental radiology

  • Christopher Walker

      Affiliations

    • Physics Department, St Lukes Hospital, Rathgar, Dublin 6, Ireland
    • Corresponding Author InformationCorresponding author.
  • ,
  • Wil van der Putten

      Affiliations

    • Physics Department, St Lukes Hospital, Rathgar, Dublin 6, Ireland
    • School of Physics, National University of Ireland, Galway, Ireland

Received 27 August 2010; received in revised form 27 November 2010; accepted 13 December 2010. published online 10 January 2011.

Abstract 

Diagnostic Reference Levels provide a method of ensuring that patient doses in medical procedures are kept at acceptable levels. Their application in dentistry can provide an indication of current dose levels and can assist in potentially significant dose reduction in Ireland given the number of patients screened annually.

This study involved retrospective analyses of entrance surface dose and dose-width-product measurements obtained in Irish Dental Practices for both Intra-Oral and Panoramic units respectively, followed by comparisons with Monte-Carlo generated computer models of these procedures. Analysis was performed on data from 33 Intra-Oral units for an Adult Mandibular Molar entrance surface dose, 198 readings for a proposed mGy/mAs reference level and 50 Panoramic machines for a dose-width product investigation.

The third quartile value of the entrance surface dose for a standard Adult Mandibular Molar Intra-Oral radiograph is (2.40 ± 0.92)mGy, compared to a computer-modelled value of 2.60 mGy. The third quartile mGy/mAs value for Intra-Oral procedures is (1.03 ± 0.38)mGy/mAs, compared to a computer-modelled value of 0.75 mGy/mAs. The third quartile dose width product for an Adult Panoramic radiograph is (59.89 ± 20.97)mGymm, compared to a computer-modeled value of 62.40 mGymm.

It is proposed to introduce Diagnostic Reference Levels of 2.4 mGy for an Adult Mandibular Molar Intra-Oral radiograph and 60 mGymm for an Adult Panoramic radiograph. The use of a new reference quantity in Intra-Oral radiology is also suggested. This has a value of 1 mGy/mAs and may be introduced alongside established procedures. These levels can be taken as guides to acceptable doses, but it should be noted that further reductions are practical under ALARA principles.

Keywords: Diagnostic Reference Levels, Dental X-ray units, Quality Assurance, Monte-Carlo, Intra-Oral, Pan-oral tomograph

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1120-1797(10)00079-7

doi:10.1016/j.ejmp.2010.12.003

Physica Medica: European Journal of Medical Physics
Volume 28, Issue 1 , Pages 7-12, January 2012